空气灭菌站在消化内镜中心空气质量控制中的应用
作者:
作者单位:

第二军医大学长海医院消化科

作者简介:

通讯作者:

中图分类号:

基金项目:


Application of air sterilization station to air quality control of digestive endoscopy center
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 文章评论
    摘要:

    目的 通过全国多中心、动态监测研究,明确空气灭菌站在消化内镜中心空气质量控制的应用效果。方法 本研究联合全国不同地区共15家消化内镜中心共同开展,每中心选取开展4级、3级消化内镜诊疗技术的内镜诊室2间,1间放置空气灭菌站(实验组),1间不放置该设备(对照组),检测并对比使用空气灭菌站前后PM5和PM0.5浓度。结果 在不使用空气灭菌设备的情况下,所有中心内镜诊室的空气质量未能达到气洁净度8级标准。动态环境下各中心实验组在使用空气灭菌站后PM5与PM0.5浓度均低于对照组,差异有统计学意义(P<0.05);各中心实验组PM5与PM0.5浓度随开机时间呈下降趋势,且在不同开机时间差异有统计学意义(P<0.05)。空气灭菌站开机15.7 min后PM5浓度可达到空气洁净度8级标准,开机25.0 min后PM0.5浓度可达到空气洁净度8级标准。结论 在消化内镜诊室这一动态环境中,空气灭菌站能够显著改善诊室内空气质量,值得在各级医疗机构内镜诊疗室推广使用。

    Abstract:

    Objective To investigate the effect of air sterilization station on air quality control of digestive endoscopy center through a national multi-center dynamic monitoring research. Methods This study was conducted jointly with 15 digestive endoscopy centers in different regions of China. Each center selected 2 endoscopy rooms using level 4 and level 3 digestive endoscopy techniques, one with an air sterilization station (the experimental group) and the other without the device (the control group). The concentrations of PM5 and PM0.5 in the rooms before and after the use of air sterilization station were detected and compared. Results The air quality of all research units failed to reach the standard air cleanliness level 8 without using air sterilization station. After using the air sterilization station, the PM5 concentration and PM0.5 concentration of the experimental group in each center under dynamic conditions were lower than those of the control group, and the difference was significant (P<0.05). The PM5 concentration and PM0.5 concentration in the experimental group at each center showed a decreasing trend with start-up time, with significant difference at different start-up times (P<0.05). The PM5 concentration could reach the air cleanliness level 8 standard 15.7 minutes after the air sterilization station started, and the PM0.5 concentration could reach the air cleanliness level 8 standard 25.0 minutes after the air sterilization station started. Conclusion In the dynamic environment of the digestive endoscopy center, the air sterilization station can significantly improve the air quality in the endoscopy rooms, which is worth promoting in endoscopy rooms of medical institutions of all levels.

    参考文献
    相似文献
    引证文献
引用本文

高野,王天骄,王洛伟,等.空气灭菌站在消化内镜中心空气质量控制中的应用[J].中华消化内镜杂志,2023,40(12):992-996.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2022-01-24
  • 最后修改日期:2023-12-03
  • 录用日期:2022-05-06
  • 在线发布日期: 2023-12-04
  • 出版日期:
您是第位访问者

通信地址:南京市鼓楼区紫竹林3号《中华消化内镜杂志》编辑部   邮编:210003

中华消化内镜杂志 ® 2024 版权所有
技术支持:北京勤云科技发展有限公司