Efficacy and safety of endoscopic marking with SPOT for gastrointestinal lesions
Author:
Affiliation:

Department of Gastroenterology, Beijing Friendship Hospital, Capital Medical University

Fund Project:

Beijing Science and Technology Plan (Z191100006619082)

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 文章评论
    摘要:

    目的 验证一种含炭定位标记液(SPOT)临床应用于内镜下病变标记的有效性和安全性。方法 前瞻性纳入2019年4月—2019年11月期间在首都医科大学附属北京友谊医院和首都医科大学附属北京朝阳医院确诊胃肠道病变需行内镜治疗或外科手术的115例患者,内镜检查时应用SPOT标记病灶,内镜治疗或外科手术时寻找标记点。采用单组目标值法计算产品标记有效率。观察标记后有无不良事件发生,比较标记前后患者血常规和肝肾功能等变化情况。结果 受试者SPOT标记的有效率为99.13%(114/115),标记维持时间可达57 d,标记过程中无穿刺肠壁或注射到腹腔情况。标记后,仅1例患者出现轻度发热,不良事件发生率为23.48%(27/115),均与试验器械无关。标记前后患者血常规和肝肾功能检查差异均无统计学意义(P>0.05)。结论 SPOT内镜定位标记液对胃肠道病灶能有效、安全地进行染色标记,无严重不良事件发生,可满足临床使用要求。

    Abstract:

    Objective To evaluate the efficacy and safety of SPOT (GI Supply, USA), a new carbon-based permanent marker approved by the Food and Drug Administration (FDA), in the endoscopic marking for gastrointestinal lesions. Methods A total of 115 patients with gastrointestinal lesions who underwent endoscopic treatment or surgery in Beijing Friendship Hospital or Beijing Chao-Yang Hospital from April 2019 to November 2019 were enrolled in the study. SPOT was used to mark the lesions, and marking points were found during endoscopic treatment or surgery to calculate the effective marking rate by single-group target value method. Adverse events after marking were recorded, and the changes of blood routine test, liver and kidney functions before and after marking were compared. Results The effective rate of endoscopic marking with SPOT was 99.13% (114/115). The longest marking time was 57 days. There was no puncture of intestinal wall or injection into abdominal cavity during the marking process. One patient developed mild fever after marking. The incidence of adverse events was 23.48% (27/115), which were all unrelated to the test equipment. There was no significant difference in blood routine tests or liver and kidney functions before and after marking (P>0.05). Conclusion SPOT produced by GI Supply can effectively mark gastrointestinal lesions without serious adverse events, which meets the requirements of clinical use.

    参考文献
    相似文献
    引证文献
引用本文

王佳旭,武珊珊,王文海,等.内镜定位标记液用于内镜下病变标记的有效性和安全性临床研究[J].中华消化内镜杂志,2022,39(9):701-706.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2021-08-01
  • 最后修改日期:2022-08-18
  • 录用日期:2021-11-15
  • 在线发布日期: 2022-08-22
  • 出版日期: 2022-09-20
您是第位访问者

通信地址:南京市鼓楼区紫竹林3号《中华消化内镜杂志》编辑部   邮编:210003

中华消化内镜杂志 ® 2024 版权所有
技术支持:北京勤云科技发展有限公司